  HTP501 Datasheet HTML 1Page - E+E Elektronik Ges.m.b.H HTP501 Datasheet HTML 2Page - E+E Elektronik Ges.m.b.H HTP501 Datasheet HTML 3Page - E+E Elektronik Ges.m.b.H HTP501 Datasheet HTML 4Page - E+E Elektronik Ges.m.b.H HTP501 Datasheet HTML 5Page - E+E Elektronik Ges.m.b.H HTP501 Datasheet HTML 6Page - E+E Elektronik Ges.m.b.H HTP501 Datasheet HTML 7Page - E+E Elektronik Ges.m.b.H HTP501 Datasheet HTML 8Page - E+E Elektronik Ges.m.b.H  
Zoom Inzoom in Zoom Outzoom out
 5 / 8 page
background image
Datasheet HTP501
v1.2 / Modification rights reserved | 5
Technical Data
Measurands
Relative humidity (RH)
Measuring range
0...100 %RH
Accuracy1)
-15...+40 °C (5…104 °F)
-15...+40 °C (5…104 °F)
-25...+70 °C (-13…+158 °F)
-40...120 °C (-40…+248 °F)
(RH ≤ 90 %)
±(1.3 + 0.003*mv) %RH
(RH > 90 %)
±2.3 RH
±(1.4 + 0.01*mv) %RH
±(1.5 + 0.015*mv) %RH
mv = measured value
Response time t90
@ 20 °C (68 °F)
<15 s
1) Including hysteresis, non-linearity and repeatability
Traceable to international standards, administrated by NIST, PTB, BEV...
The accuracy statement includes the uncertainty of the factory calibration with an enhancement factor k=2 (2-times standard deviation).
The accuracy was calculated in accordance with EA-4/02 and with regard to GUM (Guide to the Expression of Uncertainty in Measurement).
Temperature (T)
Measuring range
-40...+120 °C (- 40...+248 °F)
Accuracy1)
-40 -20
0
20 40 60 80 100 120 140 160 180
0
0.1
0.2
0.3
0.4
0.5
0.6
∆
T [°C]
T [°C]
40 20
0
40
20
80
60
120
100
0.15
0.45
0.1
0.2
0.3
0.4
0.5
0
∆T [°C]
T [°C]
± ∆T [°C]
T [°C]
± ∆T [°C]
T [°C]
1) Traceable to international standards, administrated by NIST, PTB, BEV...
The accuracy statement includes the uncertainty of the factory calibration with an enhancement factor k=2 (2-times standard deviation).
The accuracy was calculated in accordance with EA-4/02 and with regard to GUM (Guide to the Expression of Uncertainty in Measurement).
Values in mm (inch)
Dimensions
code “cable length”
code “probe length”
32
(1.3)
15
(0.6)
45 (1.8)
16
(0.6)



Html Pages

1 2 3 4 5 6 7 8


데이터시트 다운로드

Go To PDF Page


링크 URL



ALLDATASHEET 가 귀하에 도움이 되셨나요?  [ DONATE ] 

Alldatasheet는?   |   광고문의   |   운영자에게 연락하기   |   개인정보취급방침   |   링크 투 데이터시트    |   링크교환   |   제조사별 검색
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com