Back Button HV7358 Datasheet HTML 20Page - Microchip Technology HV7358 Datasheet HTML 21Page - Microchip Technology HV7358 Datasheet HTML 22Page - Microchip Technology HV7358 Datasheet HTML 23Page - Microchip Technology HV7358 Datasheet HTML 24Page - Microchip Technology HV7358 Datasheet HTML 25Page - Microchip Technology HV7358 Datasheet HTML 26Page - Microchip Technology HV7358 Datasheet HTML 27Page - Microchip Technology HV7358 Datasheet HTML 28Page - Microchip Technology Next Button
Zoom Inzoom in Zoom Outzoom out
 24 / 50 page
background image
HV7358
DS20005918A-page 24
 2018 Microchip Technology Inc.
4.4.1
BROADCASTING vs. DAISY-CHAIN
MODE
The fast programming interface can also operate in two
modes: Broadcasting mode or Daisy-Chain mode. The
pin, SPIB, selects the operating mode. When the SPIB
pin is high, the SPI interface operates in Broadcasting
mode. When SPIB = 0, it operates in Daisy-Chain
mode. Alternatively, the operating mode can be
selected via the SPIBC bit in the I2C register, to control
the SPI mode, if the override bit, SPISEL = 1.
In Daisy-Chain mode, the LVDS output SDO is the Shift
register output. SDO is the fast SPI Shift register out-
put, which is either the 12-bit Delay register DLY[11:0]
or 26-bit DLY[11:0], TLP[6:0] and TLN[6:0], depending
on the SDLY selection. Data is clocked out of the Trans-
mit Pattern Shift register onto SDO on the rising edge
of the CKOP.
In Broadcasting mode (pin SPIB = 1), the LVDS
outputs, CSOP/CSON and SDOP/SDON, are matched
in delay with the CKOP/CLON in timing, as shown in
Figure 4-16.
The Broadcasting mode should be used when all
devices need to be programmed with the same data. In
the medical ultrasound application, all the transmit
channels have the same transmit patterns. In that
scenario, the Broadcasting mode should be used to
reduce the programming time. The Daisy-Chain mode
should be used when the programming data is different
across devices. In most medical ultrasound applica-
tions, the transmit delay is different from channel to
channel. In that scenario, the Daisy-Chain mode
should be used when programming the channel delay
counters.
Note that the programming clocks, CLKP/N, are only
used for programming the transmit patterns and the
channel delay counters. The CLKP/N do not need to be
active all the time and they can be turned off after SPI
writing.
FIGURE 4-2:
Connection Circuit Block Diagram of Fast SPI Interface.
HV7358
U2
U4
LVDS
SDI
TXRW
Shift Reg.
SPIB
D0
Qn
CLK
CS
SDO
CSO
TXRW
CLK
CS
SDI
FPGA
SDO
U1
HV7358
HV7358
SPIB
SDI
TXRW
Shift Reg.
SPIB
D0
Qn
CLK
CS
SDO
CSO
LVDS
SDI
TXRW
Shift Reg.
SPIB
D0
Qn
CLK
CS
SDO
CSO
CKO
CKO
CKO



Html Pages

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50


데이터시트 다운로드

Go To PDF Page


링크 URL



ALLDATASHEET 가 귀하에 도움이 되셨나요?  [ DONATE ] 

Alldatasheet는?   |   광고문의   |   운영자에게 연락하기   |   개인정보취급방침   |   링크 투 데이터시트    |   링크교환   |   제조사별 검색
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com