Back Button SDP32 Datasheet HTML 2Page - Sensirion AG Switzerland SDP32 Datasheet HTML 3Page - Sensirion AG Switzerland SDP32 Datasheet HTML 4Page - Sensirion AG Switzerland SDP32 Datasheet HTML 5Page - Sensirion AG Switzerland SDP32 Datasheet HTML 6Page - Sensirion AG Switzerland SDP32 Datasheet HTML 7Page - Sensirion AG Switzerland SDP32 Datasheet HTML 8Page - Sensirion AG Switzerland SDP32 Datasheet HTML 9Page - Sensirion AG Switzerland SDP32 Datasheet HTML 10Page - Sensirion AG Switzerland Next Button
Zoom Inzoom in Zoom Outzoom out
 6 / 14 page
background image
www.sensirion.com
Version 0.8 – January 2017
6/14
5. Digital Interface Description
The SDP3x-digital interface is compatible with the I2C protocol. This chapter describes the command set for SDP3x-digital.
For detailed information about the I2C protocol, please check the document "NXP I2C-bus specification and user manual".
5.1
I2C Address
The I2C address for SDP3x-digital can be selected with the ADDR pin. For more information refer to section 3.3.
5.2
I2C Sequences
The commands are 16-bit. Data is read from the sensor in multiples of 16-bit words, each followed by an 8-bit checksum to
ensure communication reliability.
I2C master writes 16 bit command
I2C master sends read header and receives multiple 16bit words with CRC byte.
Dark areas with white text indicate that the sensor controls the SDA (Data) line.
I2C sequences can be aborted with a NACK and STOP condition.
5.3
I2C Commands
The command set of the SDP3x-digital consists of a set of different commands:
-
Continuous measurement
o
Start Continuous measurement commands
o
Stop measurement command
-
Triggered measurement commands
-
Read product identifier and serial number
-
Soft reset
S
W
I2CAdr[6:0]
Cmd[15:8]
Cmd[7:0]
S
R
I2CAdr[6:0]
Data1[15:8]
Data1[7:0]
CRC1[7:0]
P
Data2[15:8]
Data2[7:0]
CRC2[7:0]
P
or
DataX[15:8]
DataX[7:0]
CRCX[7:0]
P
or
≈



Html Pages

1 2 3 4 5 6 7 8 9 10 11 12 13 14


데이터시트 다운로드

Go To PDF Page


링크 URL



ALLDATASHEET 가 귀하에 도움이 되셨나요?  [ DONATE ] 

Alldatasheet는?   |   광고문의   |   운영자에게 연락하기   |   개인정보취급방침   |   링크 투 데이터시트    |   링크교환   |   제조사별 검색
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com