| 전자부품 데이터시트 검색엔진 |
|
ADPD4100 데이터시트(PDF) 44 Page - Analog Devices |
|
|
|||||||||||||||||||||||||||||
ADPD4100 데이터시트(HTML) 44 Page - Analog Devices |
|
44 / 101 page ![]() ADPD4100/ADPD4101 Data Sheet Rev. 0 | Page 44 of 101 ECG MEASUREMENT WITH THE ADPD4100/ADPD4101 The ADPD4100/ADPD4101 can be used for ECG applications with the simple addition of an external RC network consisting of two 500 kΩ resistors in series with the inputs and a 470 pF capacitor across the inputs, as shown in Figure 52. The electrical equivalent model for an electrode is shown together with the RC circuit, external to the ADPD4100/ADPD4101. The 500 kΩ resistors limit the current that can be injected into or pulled from the body in the case of shorted input pins. The 470 pF capacitor serves as the sensing capacitor for the ECG signal. The ECG signal is integrated onto the sensing capacitor. The value of this capacitor is chosen such that an acceptable SNR is achieved and the time constant of the RC network and the electrode skin contact allows sufficient charge accumulation on the sensing capacitor during the sampling period. The RC network of the 500 kΩ resistors and the 470 pF capacitor also acts as a low-pass filter, which helps reduce the high frequency noise due to electrode skin contact. For multilead ECG measurement, each lead requires a separate pair of inputs and the RC network of two 500 kΩ resistors and the 470 pF capacitor, as shown in Figure 52. TIA RF RF RIN RIN 500kΩ 500kΩ 470pF RP CP RP CP ADPD4100/ ADPD4101 INx INx VDC VDC RS RS RA LA SKIN/ELECTRODE INTERFACE NOTES 1. RA is right arm. LA is left arm. Figure 52. Circuit for Measuring Single-Lead ECG with the ADPD4100/ADPD4101 Sleep Float Mode The ADPD4100/ADPD4101 ECG measurement operates in sleep float mode. Sleep float mode allows a robust ECG measurement regardless of whether low impedance wet electrodes or high impedance dry electrodes are used. In sleep float mode, the sensing capacitor floats all the time except the charge transfer, accumulating charge from the ECG signal. The accumulated charge is then transferred into the ADPD4100/ADPD4101 during a specified time slot for charge measurement. The device must be configured to float the inputs for the ECG during the preconditioning period and during sleep. The inputs are connected to the external capacitor only during the charge transfer. At all other times, the inputs for the ECG are floating, resulting in a float time of 1/tP, where tP is the sampling rate of the ADPD4100/ADPD4101. For example, the float time in sleep float mode is ~3.3 ms for the sampling rate of 300 Hz. An advantage of using sleep float in ECG measurements is the longer charging time for the sensing capacitor. In sleep float mode, charge accumulation on the sensing capacitor happens during the sleep and during all other enabled time slots. The time slot associated with the sleep float mode is only used to transfer the charge from the sensing capacitor to the ADPD4100/ ADPD4101 amplifier. Sleep float mode also allows the use of other time slots for different sensor-based applications while ECG is being measured because the sensing capacitor floats regardless of the types of applications that the other time slots enable. Another advantage of using sleep float mode in ECG measure- ments is the reduced power consumption and noise. In sleep float mode, while the sensing capacitor is floating, it is discon- nected from the amplifier and the amplifier is not powered. Therefore, sleep float mode reduces the power consumption by using the passive charging time to transfer the ECG signal from the body onto the sensing capacitor while the amplifier and all the later stages are powered down. The passive charging process of the sampling capacitor associated with sleep float mode also reduces the noise as the charging process is noise free. A timing diagram for sleep float mode is shown in Figure 53. The relevant register settings for sleep float mode are shown in Table 27. INTEGRATOR OUTPUT INTEGRATOR SEQUENCE ADC READ ACCUMULATED CHARGE ON CAPACITOR CONNECT/FLOAT WAKE TIME SLOT A (TS A) SLEEP TS B TS X SEQUENCE TIMING MOD_WIDTH_x WAKE MOD_OFFSET_x INTEG_OFFSET_x INTEG_WIDTH_x START OF TIME SLOT Figure 53. Sleep Float Mode Timing Diagram |
|
링크 URL |
| ALLDATASHEET 가 귀하에 도움이 되셨나요? [ DONATE ] |
Alldatasheet는? | 광고문의 | 운영자에게 연락하기 | 개인정보취급방침 | 링크 투 데이터시트 | 링크교환 | 제조사별 검색 All Rights Reserved©Alldatasheet.com |
| Russian : Alldatasheetru.com | Korean : Alldatasheet.co.kr | Spanish : Alldatasheet.es | French : Alldatasheet.fr | Italian : Alldatasheetit.com Portuguese : Alldatasheetpt.com | Polish : Alldatasheet.pl | Vietnamese : Alldatasheet.vn Indian : Alldatasheet.in | Mexican : Alldatasheet.com.mx | British : Alldatasheet.co.uk | New Zealand : Alldatasheet.co.nz |
|
Family Site : ic2ic.com |
icmetro.com |