| 전자부품 데이터시트 검색엔진 |
|
ADPD4100 데이터시트(PDF) 47 Page - Analog Devices |
|
|
|||||||||||||||||||||||||||||
ADPD4100 데이터시트(HTML) 47 Page - Analog Devices |
|
47 / 101 page ![]() Data Sheet ADPD4100/ADPD4101 Rev. 0 | Page 47 of 101 Improving SNR with Integrator Chopping in ECG Measurements Integrator chopping can also improve SNR in ECG measurements with the sleep float mode by eliminating low frequency noise content. The procedure to enable integrator chop mode is the same as explained previously in the Improving SNR Using Integrator Chopping section. However, ECG measurements with integrator chopping require additional considerations. The sign of the dc offset voltages in Figure 52 can be positive or negative. When the sign of the net dc offset voltage is negative, integrator chop mode can result in clipping of the ECG signal. To prevent clipping of the ECG signal, lit data must be used. Figure 23 shows the datapath. In Figure 23, lit and dark values can be optionally written to FIFO. This option allows the use of negative signal values by writing both lit and dark values to the FIFO, and the user can perform a signed subtract in external processing of the data to calculate the signal value. Therefore, integrator chopping can be used regardless of the sign of net dc offset voltage. Lead Off Detection To perform a lead off detection measurement, the ADPD4100/ ADPD4101 measure the impedance of the electrode skin contacts to determine whether one or more of the electrodes are not making contact with the skin. Lead off measurement can be performed in two different ways based on the number of electrodes used. Three-Electrode Lead Off Measurement The three-electrode configuration requires a third electrode connected to an unused VCx pin to provide a stimulus to the body. The RC network of the ECG measurement is bypassed by wiring the electrodes directly to a separate set of inputs through 25 kΩ resistors. The response from the stimulus is measured from this separate set of inputs. Three-electrode lead off measurement is capable of determining which electrode is loose or has lost the contact with the skin. Figure 54 shows the circuit for the three-electrode lead off detection measurement. RBODY is the resistance of the body. CH2 CH1 IN1 IN2 IN4 470pF VC2 VC2 ADPD4100/ ADPD4101 500kΩ 500kΩ 25kΩ 25kΩ IN3 25kΩ RBODY RBODY RBODY E1 E2 E3 47pF 47pF 47pF Figure 54. Circuit Used for Three-Electrode Lead Off Detection Measurement ECG and three-electrode lead off detection are measured as follows: 1. ECG is measured in Time Slot A as defined in the ECG Measurement with the ADPD4100/ADPD4101 section. 2. Lead off detection of the ECG electrodes is taken in Time Slot B by making simultaneous single-ended impedance measurements of ECG Electrode 1 (E1) and Electrode 2 (E2) into Channel 1 and Channel 2, respectively. When both ECG electrodes, E1 and E2, are making contact with the skin during the measurement, an ECG signal is visible. The impedance measurements of the E1 and E2 electrodes have some readout indicating that contact is being made with the skin and current is flowing into the ADPD4100/ADPD4101 through the body of low impedance when the stimulus is applied. When either ECG electrode stops making contact with the skin or is loose, there is no ECG signal in the acquired trace. When contact between both electrodes and the skin is restored, the ECG signal appears immediately. Because two inputs and two channels are allocated to detect leadoff condition for two electrodes, this measurement method can determine if one electrode loses contact with skin or both electrodes lose contact. This measurement can also detect which one of the electrodes loses contact with skin if only one electrode loses contact. Figure 55 illustrates a representation of ADC output changes in different cases for leadoff condition. In Figure 55, before Time tA, both ECG electrodes make contact with the skin. At Time tA, E1 is disconnected from the skin. The time between tA and tB shows the case where only E1 is disconnected from the skin. At Time tB, E1 starts to make contact with skin and output of the two channels go to their initial levels. At Time tC, only E2 is disconnected from the skin and it stays disconnected until Time tD. At Time tE, E2 starts making contact with the skin again. At Time tE, both E1 and E2 are disconnected from the skin, and they stay disconnected until Time tF. At Time tF, both E1 and E2 start making contact with the skin. Therefore, lead off condition is detected in all electrode connection cases, and three-electrode lead off measurement detects and distinguishes all the different cases. The level of actual ADC outputs associated with Channel 1 and Channel 2 may differ because the type and placement of the electrodes may be different in each case, and RBODY differs from person to person, which affects the amount of current that each channel receives. TIME tA tB tC tD tE tF CHANNEL 2 CHANNEL 1 Figure 55. Graph of Three-Electrode Lead Off Measurement |
|
링크 URL |
| ALLDATASHEET 가 귀하에 도움이 되셨나요? [ DONATE ] |
Alldatasheet는? | 광고문의 | 운영자에게 연락하기 | 개인정보취급방침 | 링크 투 데이터시트 | 링크교환 | 제조사별 검색 All Rights Reserved©Alldatasheet.com |
| Russian : Alldatasheetru.com | Korean : Alldatasheet.co.kr | Spanish : Alldatasheet.es | French : Alldatasheet.fr | Italian : Alldatasheetit.com Portuguese : Alldatasheetpt.com | Polish : Alldatasheet.pl | Vietnamese : Alldatasheet.vn Indian : Alldatasheet.in | Mexican : Alldatasheet.com.mx | British : Alldatasheet.co.uk | New Zealand : Alldatasheet.co.nz |
|
Family Site : ic2ic.com |
icmetro.com |